8.2.15 Simultaneous Glucose and L-Lactate
Refer to the sections above on Glucose and L-Lactate for theoretical and special considerations. Then follow the
instructions below to set up your instrument for simultaneous determination.
A Probe
B Probe
Chemistry
Glucose
L-Lactate
System Buffer
YSI 2357
Calibrator Std
YSI 2776
Linearity Std
YSI 1531
YSI 1530
Membrane
YSI 2365
YSI 2329
Sample Size
25 µL
Unit of Conc.
g/L
Cal Value
2.50
0.50
End Point
30 sec
Note: See Appendix BConcentration Unit Conversion if concentration unit conversion is required
.