140
11.1.2 Ion Selective Electrodes
Listed below are example printouts and explanations of the information for ISE sensors.
Calibration Report (Brief) Calibration Report (Detail)
==Calibration Report==
Probe3A
500 mg/L NH4+ 2972
NH4+
K+
-----
-----
IB(mV):
28.01
26.97
NPL(mV):
24.31
1.43
IB(mV):
28.09
27.03
NPL(mV):
24.36
1.65
IB(mV):
28.10
27.06
NPL(mV):
24.30
1.31
Avg:
24.32
1.46
Std Dv:
0.03
0.17
End Point
30 sec
Sample size
25 uL
Temperature
21.50 C
Probe3B
1000 mg/L K+ 2971
NH4+
K+
-----
-----
IB(mV):
28.40
26.98
NPL(mV):
6.97
37.42
IB(mV):
28.50
26.97
NPL(mV):
6.31
37.45
IB(mV):
28.63
27.02
NPL(mV):
6.09
37.42
Avg:
6.12
37.43
Std Dv:
0.17
0.02
End Point
30 sec
Sample size
25 uL
Temperature
21.50 C
---------------
->Coef:
0.19
0.02
->Slp:
57.18
60.05
---------------
Tue 10/25/2015 11:08:32
YSI 2950 13J000024
==========================
==Calibration Report==
Probe3A
500 mg/L NH4+ 2972
NH4+
K+
-----
-----
Avg:
24.32
1.46
Std Dv:
0.03
0.17
Probe3B
1000 mg/L K+ 2971
NH4+
K+
-----
-----
Avg:
6.12
37.43
Std Dv:
0.17
0.02
---------------
->Coef:
0.19
0.02
->Slp:
57.18
60.05
---------------
Tue 10/25/2015 11:08:32
==========================