66
Touch the Setpoint [10.00g/L] button and enter the threshold value.
Set the Error Direction to trigger when the measured analyte value goes [Under] the threshold or [Over] the threshold.
Touch the Feed Pump Line [None] button and select the 2960 output line that your external alarm/pump is connected to.
The default output is Active High (+5V). Touch the Feed Pump Line [High] button to change it to [Low] if your alarm/pump
requires an Active Low signal (0V) to turn on.
Touch [Save] to save changes and return to the Edit Chamber 1 screen.
Touch [Save] to save changes and return to the Edit Configuration screen.