4
1
SPECIFICATIONS
Power SupplySpecifications
Power supply Operating Condition Units
Voltage Range 100-240 VoltsAC
Frequency 50/60 Hertz
Current 0.6 Amps
Environmental OperatingConditions
Environmental Condition Operating Condition Units
Altitude 3000(max) Meters
Operating Temperature 16– 40 ºC
Storage Temperature -10to 50 ºC
Operating Relative Humidity 20-80(non-condensing) %
Voltage Range 12 VoltsDC
Frequency DC
Current 1.2 Amps
Installation/Over Voltage CategoryII
Pollution Degree 2
INSTRUCTIONS FOR USE – InitialSet-up
ChooseLocation
Place 3M™Attest™ Auto-reader390on a firm level surface away from sunlight and
incandescent light. To save counter space, units may be stacked twohigh.
WARNING:
Do not place the instrument in environment exposed to sunlight or strong
incandescentlight.
Do not place instrument close to any device that emits a strong
electro-magneticfield.
Do not use on a vibrating surface.
Select & Install Correct PowerPlug
The 3M™Attest™ Auto-reader390power supply comes with multiple blade
configurations. Install the blade configuration appropriate for your country into the power
supply and plug into appropriate AC groundedoutlet.
WARNING: Use only the power supply specified for this product and certified for
the country of use.
Wait 30Minutes BeforeUsing
Once plugged in, allow a 30minute warm-up period before placing 3M™Attest™ Rapid Readout Biological Indicators into the incubation wells. The C1caution code will disappear
when proper incubation temperature isreached.
SpecialConsiderations
It is recommended that the unit be left on to eliminate warm-up periods. The unit may be turned off when it is anticipated that it will not be used for an extendedperiod.