Setting the Incubation WellConfiguration
Incubation wells 1-5and 6-10can be configured to run the same or different types of 
3M™Attest™ Rapid Readout Biological Indicators. The flashing “A” in Attest indicates the 
setting may be changed by the user. There are three incubation well configurationoptions:
LCD Panel Configuration Setting Result
Attest 1291 Attest 1292
Auto-reader wells 1- 5configured for 
1291BIs (blue cap, 1hour readout) and wells 
6- 10configured for 1292BIs (brown cap, 
3hour readout)
Attest 1291 Attest 1291
Auto-reader wells 1–10configured for 
1291BIs (blue cap, 1hour readout)
Attest 1292 Attest 1292
Auto-reader wells 1–10configured for 
1292BIs (brown cap, 3hour readout)
Use the Up Arrow or Down Arrow buttons to move through the three options until the 
desired configuration setting isdisplayed.
Press 
 to select the desired option and advance to the Acceptscreen.
Note 1: If there are any incubating BIs which do not yet have a final result (+ or -) 
displayed, the well configuration for the occupied wells will NOT change until the final 
result is displayed on the LCD panel and the BI is removed from thereader.
Note 2: Ensure the color-coded incubation well configuration sticker surrounding the 
incubation wells matches the programmed wellconfiguration.
AcceptScreen
Use the up or down buttons to toggle between Yes and No, press   to accept changes 
(Yes selected) or retain current settings (No selected) and exit the configuration mode. The 
LCD panel will return to the standarddisplay.
INSTRUCTIONS FOR USE - ReadoutProcedure
Instructions for Biological Indicator (BI) Activation andIncubation
 WARNING: Match the cap color of the 3M™Attest™ Rapid Readout Biological 
Indicator with the color-coded configuration sticker surrounding the 3M™Attest™ 
Auto-reader390incubation wells.
 CAUTION: Wear safety glasses when activating 3M™Attest™ Rapid Readout 
Biological Indicators.
 CAUTION: Allow the 3M™Attest™ Rapid Readout Biological Indicator to cool 
for the recommended time period before activating. Activating or excessive handling of 
the biological indicator before cooling may cause the glass ampoule to burst.
 CAUTION: To avoid the possibility of the 3M™Attest™ Rapid Readout 
Biological Indicator vial absorbing fluorescent residue from a chemical indicator or 
tape, place the 3M™Attest™ Rapid Readout Biological Indicator vial so it does not 
come in direct contact with chemical indicators or tape.
Follow this procedure for every 3M™Attest™ Rapid Readout Biological Indicator to 
beread.
While wearing safety glasses, CLOSE the cap of 
the cooled 3M™Attest™ Rapid Readout Biological 
Indicator (BI) by pushing capdown.
CRUSH the glass media ampoule in the Crusher Well 
built into the top of the 390Auto-reader.
Holding BI by the cap, TAP the bottom of the vial on a 
tabletop until media wets spore strip at bottom of vial. 
Do not tap BI onAuto-reader.
PLACE the crushed BI in color-coded incubation well. 
Observe 3M™Attest™ Auto-reader390LCD panel to 
confirm correct placement of the BI. If the BI is correctly 
seated, the 3M™Attest™ Auto-reader390LCD panel 
will display ***, **, * and then remaining minutes of 
incubation directly below the incubatingBI.
RECORD BIresult.
 WARNING: Do not remove the 3M™Attest™ Rapid Readout Biological 
Indicator from the incubation well until the (+) or (-) symbol on the LCD panel indicates 
the test is complete.
INSTRUCTIONS FOR USE – AdditionalInformation
BIPlacement
Wells must be configured and color coded to match the cap of the 3M™Attest™ 
Rapid Readout Biological Indicator, e.g. the 3M™Attest™ 1291Rapid Readout 
Biological Indicator with a blue cap goes into a blue well and the 3M™Attest™ 
1292Rapid Readout Biological Indicator with a brown cap goes into a brownwell.
Remaining incubation time, in minutes, will be displayed on the LCD display to indicate 
the 3M™Attest™ Rapid Readout Biological Indicator is incubating in the reader well and 
should not be touched. The time remaining display indicates the results are not available 
and incubation is inprogress.
Do not remove 3M™Attest™ Rapid Readout Biological Indicator until the (+) or (-) 
symbol on the LCD panel indicates the test iscomplete.
Note: If 3M™Attest™ Rapid Readout Biological Indicator is inadvertently removed 
before the incubation period is complete, an alarm will sound and a C2Caution code will 
be displayed on the LCD panel under that well. You must return the 3M™Attest™ 
Rapid Readout Biological Indicator to the reader well within 10seconds or the 
results will belost.
 WARNING: Do not remove or change placement of 3M™Attest™ Rapid 
Readout Biological Indicator once it is placed into a well. Doing so may result in loss of 
results and the 3M™Attest™ Rapid Readout Biological Indicator test may be invalid.