USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
4 Initialize and Run the Analyzer
To initialize the analyzer:
1. Press the power switch on the front of the analyzer to theON position. (0 = OFF / -
= ON) The internal computer initializes, and a screen (Figure 20) displays as the
program loads. The
Launch Service
screen displays after initialization. (Figure 22)
2. Click on the launch button to manually launch the analyzer.
a. The launch button is the abbreviated name of the gas analyzer. For
example, the GLA351-N2OCM button displays N2O-CO. (Figure 22)
b. If you don’t make a selection within 120-seconds, the analyzer automatically
defaults to the
Main Panel Numeric
display. (Figure 28)
3. Click on the maintenance SERVICE button (Figure 22) if you need more time or need
to choose a
maintenance setting
. (Figure 23)
Once a month, the analyzer automatically performs a thorough file
system integrity check during startup. This maintenance takes
approximately one to two minutes before it continues to load the
software.
Do not turn off the computer during the monthly maintenance!
Figure 20: Start up Screen in Busy Mode