EasyManua.ls Logo

ABB LGR GLA351 Series - Page 75

ABB LGR GLA351 Series
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
75
Integration of the measured spectra with the measured values of:
Gas temperature
Gas pressure
Path length
Line strength of the probed transition
Enables you to determine the mole fraction directly from the relation:
Use this equation to determine gas concentrations, even in hostile environments without
using calibration gases or reference standards. These values are measured:
Mixtures containing several species
Flows at elevated temperatures and pressures

Table of Contents