AS800 PRESSURE TRANSMITTER
CREATE/MODIFY CONFIGURATION
P - 3
Type a descriptive message using up to 32 charac-
ters. This field can be used to note anything of impor-
tance to the device or installation.
To select a character, press the key that has the
desired character. Continue to press the key until the
desired character appears. Use the right arrow key to
move to the next character. Use the left arrow key to
go back to the previous character.
Type a descriptor string using up to 16 characters.
This field can be used for notations about the device
or process.
To select a character, press the key that has the
desired character. Continue to press the key until the
desired character appears. Use the right arrow key to
move to the next character. Use the left arrow key to
go back to the previous character.
Enter a day and press . Enter a month and
press . Enter a year and press .
This date can represent the creation date of the con-
figuration, the date the device or element was
installed, or some other significant date.
Select LINEAR.
NOTE: Other output type selections are SQUARE
ROOT, FUNC GENERATOR and 3/2 FLOW MODE.
NOTE: IBIS software is needed to configure FUNC
GENERATOR breakpoints.
Use
to return to a previous configuration
screen from any screen in the configuration process.
Enter a value between 0 and 60 seconds.
Select an engineering unit best suited for the
application. Other units not shown include BARS,
mBAR, gSqCm, KgCm2, Pa, KPa, torr and ATM.
Input lower range value using the arrow keys, then
press . Input the upper range value, then
press .
Key Display Comments
ENTER
MESSAGE:
←PREVIOUS NEXT→
ENTER
DESCRIPTOR:
←PREVIOUS NEXT→
ENTER
DATE:
DAY: nn
MONTH: nn
YEAR: nn
ENTER
ENTER ENTER
ENTER
OUTPUT TYPE
→ LINEAR
SQUARE ROOT
FUNC GENERATOR
3/2 FLOW MODE
BACK
ENTER
DAMPING:
(0 - 60 SEC)
0.5 SECS
ENTER
ENGINEERING UNIT
→ iH20 iHg
ftH2O mmH2O
mmHg PSI
ENTER
LOWER RANGE VAL
nn.nn UNITS
UPPER RANGE VAL
nn.nn UNITS
ENTER
ENTER