PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
24
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
4. The Outline page is the default view.
Figure 1: Assay Configuration Screen, Outline Page
5. Verify Quantitative Ranges is selected (). This is the
default setting.
6. Select the Calibration tab to access the Calibration page.
Figure 2: Assay Configuration Screen, Calibration Page
7. Verify the Factor option is selected in the Calib Mode field.
8. Verify 1000 is entered in the Factor field.
9. Select <OK> in the upper-right corner of the Assay
Configuration screen. The Save dialog window displays.
10. Select <OK> to save the settings.
11. Repeat steps 1 through 10 for each DOA assay installed.
Select
Factor