PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution Flowcharts
AEROSET
®
System Troubleshooting Guide 249
94816-107—November 2004
2. If issue is only on one line, troubleshoot:
Sample probes for damage, obstruction, and
alignment
Reagent probes for damage, obstruction, and
alignment
All probe tubing for crimps or leaks
Sample, reagent, and wash solution syringes
•Mixers
•Maintenance
•Probe/mixer wash flow