AEROSET
®
System Troubleshooting Guide 255
94816-107—November 2004
QC Out of Range, Continued
1. Check for the following:
Involved assays on same line?
Involved assays using similar wavelengths?
Is the reagent kit make up the same for all assays
involved? R1 vs. R2?
2. Multiple assays? Check for the following:
Test file parameters are correct
Recent QC or reagent lot changes
Recent shifts or trends in QC data
Is QC material assayed or unassayed?
Manufacturer claims for analyte stability
Are ranges appropriate to assess system performance?
Peer group statistics
3. Verify reagent handling:
Prepared correctly, if preparation required
Reagent loaded in correct Reagent Carousel and
position
•Expiration date
Date reagent was opened and placed onboard
Pipettes—disposable or washed?
Workload pattern
4. Check calibrators for the following:
Correct values entered
Dates opened
Placed onboard in correct positions
Length of time calibrators have been onboard
5. Verify instrument:
Check when source lamp was installed
Ensure all scheduled maintenance is current
Check for damaged probes, drips, or drops
Check cuvette washer performance
Check for instrument leaks, e.g., pumps and syringes