258
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Calibration Errors, Continued
No
Yes
No
6. Replace lamp
& recalibrate.
OK?
Service
5. Check
dispense
components. OK?
No
Adjust, replace, or clean
affected component(s)
& calibrate. OK?
Yes
Yes
Open new lot of
reagent &
recalibrate. OK?
(If already done,
proceed to #6.)
4. Open new lot
of reagent &
recalibrate. OK?
Stop
Open new lot of
calibrators &
recalibrate. OK?
Yes
No
No
Yes
No
Is cuvette washer
dripping or
overflowing cuvettes?
Yes
Replace poppet
valve or look for
obstruction in HC
waste tubing.
Resolved?
No
No Service
Yes
Yes
No
Stop
No
Open new lot of
calibrators &
recalibrate. OK?
(If already done,
proceed to #6.)
No
Yes
Yes
No
Yes
Proceed
to QC
OOR
Flowchart
Proceed
to QC
OOR
Flowchart
Proceed
to QC
OOR
Flowchart
Run QC.
OK?
Run QC.
OK?
Recalibrate.
OK?
Run QC.
OK?
Run QC.
OK?
Yes
Yes
Yes
Run QC.
OK?
Yes Stop
No
Proceed
to QC
OOR
Flowchart