User Manual General instruction for use
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
87
Note
X-MIND prime is based on a standard dentition and ascending rami shape. This
shape, based on statistical data, establishes a standard shape for the
dentomaxillofacial complex, also defining the position and the direction of the
condyles. Patient anatomy can differ significantly from the statistical model;
based on experience and competence, the user has to judge this variation.
IN ANY CASE, RADIOGRAPHY IMAGES CANNOT BE USED TO PERFORM
CALCULATIONS OF DISTANCES, ANGLES ETC. ON THE IMAGE.
Warning
The measurement of lengths on digital images depends on the specific length
calibration of the software used. It is therefore very important to check the
length calibration of the software. With TMJ exams, to obtain the measurement
of the anatomical part, taking into consideration the enlargement factor, the
length calibration factor is 100 pixels = 8 mm (in the centre of the focus layer).