6146/6156 DC Voltage/Current Generators Operation Manual
7.2.5 Calibration by Manual Operation
7-15
Calibration data save
The calibration data is just stored in the work memory
so far. If the CAL mode is turned OFF as it is, the cali-
bration data will be cleared.
To enable the calibration data, save it.
Select "SAVE" using MENU, SYS and CAL.DT, and
press ENTER.
Exiting calibration mode
Switch the CAL mode from ON to OFF using MENU,
SYS and CAL.
The CAL indicator goes OFF.