6146/6156 DC Voltage/Current Generators Operation Manual
2.2.1 Setting Source Value
2-13
For example, if is pressed when the value is -1 V, it will become "-0 V."
4. Pressing again releases the zero state.
Pressing or also releases the zero state.
In this case, specifying the reverse polarity changes the polarity.
3. Changing the source range
Change the source range by using the DOWN or UP key.
The procedure of changing the range varies depending on the count fix function ON or OFF.
The following is an example with the count fix function set to OFF as factory default.
For more information on the count fix function, refer to Section 2.2.6, “Count Fix Function.”
When the count fix function is set to OFF, the range change adjusts to synchronize the source value
before and after the change.
Example using the 6156
The cursor position remains in the digit position of the value before the change (if possible).
0
0
-
+
(3 V range)
(30 V range)
UP
(3 V range)
Cursor (blinking)
Cursor (blinking)
DOWN
Cursor (blinking)