EasyManuals Logo

Advanced Instruments Osmometer 3250 Service Manual

Advanced Instruments Osmometer 3250
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
118
Upgrade Kit - 3250/4250 Suffix D to E Advanced
Osmometer 3250/Cryoscope 4250 Service Manual
14. Press STOP to exit the SETUP menu.
15. At “Press START to Continue”, insert a
sample tube containing water or calibra-
tion standard, then press START to initi-
ate the instrument self-diagnostics.
Note: During the first start-up, you may see
“A/D Cal Failure”. This is due to the
board set powering up in an instru-
ment for the first time. This error mes-
sage may be ignored. The error mes-
sage will clear and the display will
return to “Press START to Continue”,
where you would press START again
to initiate the instrument self-diagnos-
tics.
16. Once completed, the instrument will indi-
cate “Recalibration Needed”.
17. Recalibrate as prompted by the display
and per the instructions in the Users
Guide.
Note: The results displayed during the cali-
bration may be quite different from the
calibration standards used, with nega-
tive values possible with the low cali-
brator. This is normal and should be
ignored. Once the instrument has been
fully calibrated, results observed dur-
ing future calibrations will be much
closer to the calibration standards.
18. The upgrade to Suffix E is complete.
325025PM Rev2
(3255 Service Manual)
Page 6 of 6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Osmometer 3250 and is the answer not in the manual?

Advanced Instruments Osmometer 3250 Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmometer 3250
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals