80
Motor Assembly Replacement 4D3350R/325304R Advanced
Osmometer 3250/Cryoscope 4250 Service Manual
325P350 Rev1
(4D35/3255/MK05 Service Manual)
Page 4 of 4
13. Install and tighten the two motor assem-
bly mounting screws (H) below the deck.
14. Reconnect the motor harness (G) to the
driver PCB.
15. Reconnect the harness to the connector at
the bottom of the Head Up/Down PCB.
16. Replace the instrument cover, then secure
in place using the mounting screws.