1100 Series DAD and MWD User Manual 77
Using the Detector 4
Automated Run Recovery in case of temporary communication failures
Table 14 Automated Run Recovery in case of temporary communication failures
Situation Reaction On ChemStation
All OK Run in progress - Data Analysis
Run / Rawdata
Elapsed run time is running
Spectra counter is running
Data are stored on PC and on card
LAN breaks Run in progress - Data Analysis
Run / Rawdata
Error Power Fail
Elapsed run time stops
Spectra counter stops
Data continues to be stored on card
LAN recovers Run in progress - Data Analysis
Run / Rawdata
Error Power Fail cleared
Elapsed run time continues at actual time
Spectra counter continues
Data continues to be stored ion PC and on card
ChemStation tries already to add missing data
(depends on the data load).
Stop time elapsed Run in progress - Data Analysis
Prerun / Rawdata
Elapsed run time stops
Spectra counter continues
ChemStation continues to ads missing data
Run ends Ready
Run finished
Prerun / Ready
NOTE
If the detector status window is not opened, you will realize only the Power Fail error and
the long Run In Progress information until the data is recovered from disk.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals