1100 Series DAD and MWD User Manual 77
Using the Detector 4
Automated Run Recovery in case of temporary communication failures
Table 14 Automated Run Recovery in case of temporary communication failures
Situation Reaction On ChemStation
• All OK • Run in progress - Data Analysis
• Run / Rawdata
• Elapsed run time is running
• Spectra counter is running
• Data are stored on PC and on card
• LAN breaks • Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail
• Elapsed run time stops
• Spectra counter stops
• Data continues to be stored on card
• LAN recovers • Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail cleared
• Elapsed run time continues at actual time
• Spectra counter continues
• Data continues to be stored ion PC and on card
• ChemStation tries already to add missing data
(depends on the data load).
• Stop time elapsed • Run in progress - Data Analysis
• Prerun / Rawdata
• Elapsed run time stops
• Spectra counter continues
• ChemStation continues to ads missing data
• Run ends • Ready
• Run finished
• Prerun / Ready
NOTE
If the detector status window is not opened, you will realize only the Power Fail error and
the long Run In Progress information until the data is recovered from disk.