EasyManuals Logo

Agilent Technologies 1260 Infinity User Manual

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
1260 Infinity Binary Pump User Manual 61
Using the Pump
4
Hints for Successful Use of the Binary Pump
Hints for Successful Use of the Binary Pump
Place solvent cabinet with the solvent bottles always on top (or at a
higher level) of the pump.
When using the binary pump without vacuum degasser, shortly degas
your solvents by putting the solvent to a suitable container and
applying a vacuum pressure for some time. If possible apply solvent
conditions that will decrease the gas solubility over time (for example,
warming up the solvents).
The use of a vacuum degasser is mandatory for flow rates below
0.5 mL/min and for configurations without damper and mixer.
When using the binary pump with vacuum degasser, flush the degasser
with at least 5 mL per channel before operating the pump, especially
when the pumping system had been turned off for a certain length of
time (for example, overnight) and volatile solvent mixtures are used in
the channels (see “Regular Priming” on page 56).
Prevent blocking of solvent inlet filters (never use the pump without
solvent inlet filters). Growth of algae should be avoided (see “Prevent
Blocking of Solvent Filters” on page 76).
Check purge valve frit and column frit in regular time intervals. A
blocked purge valve frit can be identified by black, yellow or greenish
layers on its surface or by a pressure greater than 10 bar in low delay
volume configuration and 20 bar in standard configuration when
pumping distilled water at a rate of 5 mL/min with an open purge
valve.
Whenever possible use a minimum flow rate of 5 µL/min per solvent
channel to avoid crossflow of solvent into the unused pump channel.
Whenever exchanging the pump seals, the purge valve frit should be
exchanged, too.
When using buffer solutions, flush the system with water before
switching it off. The seal wash option should be used when buffer
solutions with concentrations of 0.1 M or higher are being pumped for
long periods of time.
Check the pump pistons for scratches, grooves and dents when
changing the piston seals. Damaged pistons cause micro leaks and will
decrease the lifetime of the seals.
After changing the piston seals, apply the seal wear- in procedure (see
“Seal Wear- in Procedure” on page 169).
Place the aqueous solvent on channel A and the organic solvent on
channel B. The default compressibility settings are set accordingly.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals