1260 Infinity Binary Pump User Manual 83
Optimizing Performance
5
When to Use the Low Volume Mixer
When to Use the Low Volume Mixer
The Low volume mixer ( 200 µL) (5067- 1565) is designed for use with the
1260 Infinity Binary LC System in low delay volume mode. This
configuration is typically used for 2.1 mm i.d., 1.8 µm particle size
columns, where emphasis is put on S/N ratio. The low volume mixer helps
mixing gradients starting with a low concentration of organic solvents,
which can cause noise on the baseline. The maximum benefit of the mixer
is achieved using the mixer together with FW revisions A.06.06 or higher.
The low volume mixer is installed with the corresponding holder (Low
Volume Mixer Holder (G1312- 00003)).