3 Installing the Pump
Priming the System
Priming the System
Initial Priming
When Before a degasser or solvent tubing can be used, it is necessary to prime the system. Isopropanol is
recommended as priming solvent due to its miscibility with nearly all HPLC solvents and its
excellent wetting properties.
Parts required # Description
1 Isopropanol
Preparations Connect all modules hydraulically as described in the respective module manuals.
Fill each solvent bottle with 100 mL isopropanol
Switch the system on
WARNING
When opening capillary or tube fittings, solvents may leak out.
The handling of toxic and hazardous solvents and reagents can carry health risks.
Observe appropriate safety procedures (for example, wear goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
NOTE
The purge tool of the Lab Advisor can be used to purge the pump automatically.
NOTE
If the pump is not able to draw in the solvent from the bottles, use a syringe to move the
solvent manually through tubing and degasser.
NOTE
When priming the vacuum degasser with a syringe, the solvent is drawn through the
degasser tubes very quickly. The solvent at the degasser outlet will therefore not be fully
degassed. Pump for approximately 10 minutes at your desired flow rate before starting an
analysis. This will allow the vacuum degasser to properly degas the solvent in the degasser
tubes.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals