1260 Infinity Binary Pump User Manual 55
Installing the Pump
3
Priming the System
1 Open the purge valve of the pump
2 Set the flow rate to 5 mL/min.
3 Select channel A1
4 Turn the flow on
5 Observe if the solvent in the tubing of channel A1 is advancing towards
the pump. If it isn’t, disconnect the solvent tubing from the solvent
selection valve, attach a syringe with a syringe adapter and pull the
liquid through the degasser. Reattach the tubing to the solvent selection
valve.
6 Pump 30 mL isopropanol to remove residual air bubbles.
7 Switch to the next solvent channel and repeat steps 5 and 6 until all
channels have been purged.
8 Turn the flow off and close the purge valve.