66 1260 Infinity Binary Pump User Manual
4 Using the Pump
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface
Auxiliary Pump Parameters
The auxiliary pump parameters are pre- set to fit most applications. 
Adjustments should only be made when required. Table 5 on page 66 
shows the available auxiliary parameters with their default values. 
Upon initialization, the pump ignores the Maximum Flow Gradient value. 
This can result in a rapid and uncontrolled pressure increase. 
➔ To prevent harm to the column, open the purge valve until the initialization is 
finished. 
Tabl e 5 Auxiliary pump parameters
Parameter Limits Description
• Maximum Flow 
Gradient
0.1 – 100 mL/min²
default: 
100 mL/min²
With this parameter flow rate changes can be ramped up and down slowly to 
avoid pressure shocks to the column. The default value is 100 mL/min² which 
in fact turns the function off.
• Minimum 
Stroke
20 – 100 µL
default: Auto
The volume one pump piston delivers per stroke. In general, a smaller stroke 
volume results in lower pump ripple. The Auto setting adjusts the strokes 
dynamically to the lowest possible value.
The strokes can be set individually for pump heads A and B.
• Compressibility
0 - 150·10
-6
/bar 
bar or enhanced 
compressibility 
calibration
default: use 
enhanced comp. 
calibration
For best performance, check option Use enhanced compressibility calibration. 
With this option, the pump will use solvent data libraries provided by Agilent or 
data generated by using solvent compressibility calibrations (see “Running the 
Solvent Compressibility Calibration” on page 134 ).
For details on solvent compressibility see “Binary Pump Solvent 
Compressibility Calibration” on page 133.
For backward compatibility to 400 bar pumps, the solvent compressibility can 
still be set manually for each channel when the box is unticked.