144 1260 Infinity Binary Pump User Manual
9 Diagnostic Signals
Analog Pressure Output
Analog Pressure Output
A BNC connector at the rear of the binary pump provides the reading of
the pressure sensor as analog value with a resolution of 1.33 mV/bar. The
maximum reading of 660 bar equals 800 mV. The signal is available in
real time and can be fed into an appropriate recording device (e.g.
integrator or strip chart recorder) for troubleshooting purposes.
Figure 22 Location of analog output connector
Analog pressure
output