EasyManua.ls Logo

Agilent Technologies 1260 Infinity - Index Limit

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
120 1260 Infinity Binary Pump User Manual
7 Error Information
Module Error Messages
Index Limit
Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was
too short (pump).
During initialization, the first piston is moved to the mechanical stop.
After reaching the mechanical stop, the piston reverses direction until the
encoder index position is reached. If the index position is reached too
fast, the error message is generated.
Probable cause Suggested actions
1 Irregular or sticking drive movement.
Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
2 Defective pump drive assembly.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Related product manuals