84 1260 Infinity Binary Pump User Manual
5 Optimizing Performance
When to Remove Damper and Mixer
When to Remove Damper and Mixer
The binary pump is equipped with a pressure pulsation damper and a
static mixer. The total delay volume of the pump is 600 – 800 µL
(depending on system pressure). The mixer has a volume of 400 µL.
For applications that require lowest delay volume (e.g. fast gradient
methods or gradient applications with low flow rates), damper and mixer
can be bypassed.
Figure 20 Flow Path Modifications of the Binary Pump