1260 Infinity Binary Pump User Manual 175
Maintenance
10
Installation of the Solvent Selection Valve Upgrade Kit
Installation of the Solvent Selection Valve Upgrade Kit
A solvent selection valve allows you to choose between four different
solvents that can be used with a binary pump. The valve switches between
two solvents A1 and A2 for channel A of the left pump head and two
solvents B1 and B2 for channel B of the right pump head.
When Applicable modules: This kit is compatible to the 1260 Infinity Binary Pumps G1312B and G1312C
and to the 1260 Infinity Binary Pump Clinical ed. K1312B.
Tools required Description
Screwdriver Pozidriv #1
Parts required p/n Description
G1381-60000 Solvent Selection Valve Upgrade Kit
Preparations If required, remove solvent tubes from the inlet valves.
The figures below show a Binary Pump G1312B. The kit can be used similarly for the Binary
Pump G1312C and for the Binary Pump Clinical ed. K1312B.