74 1260 Infinity Binary Pump User Manual
4 Using the Pump
Algae Growth in HPLC Systems
Algae Growth in HPLC Systems
The presence of algae in HPLC systems can cause a variety of problems
that may be incorrectly diagnosed as instrument or application problems.
Algae grow in aqueous media, preferably in a pH range of 4- 8. Their
growth is accelerated by buffers, for example phosphate or acetate. Since
algae grow through photosynthesis, light will also stimulate their growth.
Even in distilled water small- sized algae grow after some time.
Instrumental Problems Associated With Algae
Algae deposit and grow everywhere within the HPLC system causing:
Blocked solvent filters or deposits on inlet or outlet valves resulting in
unstable flow, composition or gradient problems or a complete failure of
the pump.
Small pore high pressure solvent filters, usually placed before the
injector to plug resulting in high system pressure.
PTFE frits blockage leading to increased system pressure.
Column filters to plug giving high system pressure.
Flow cell windows of detectors to become dirty resulting in higher noise
levels (since the detector is the last module in the flow path, this
problem is less common).

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals