112 1260 Infinity Binary Pump User Manual
7 Error Information
Module Error Messages
Missing Pressure Reading
Error ID: 2054
The pressure readings read by the pump ADC (analog- digital converter)
are missing.
The ADC reads the pressure signal of from the damper every 1ms. If the
readings are missing for longer than 10 s, the error message is generated.
Wrong Pump Configuration
Error ID: 2060
At switch- on, the pump has recognized a new pump configuration.
The binary pump is assigned its configuration at the factory. If the active
inlet valve and pump encoder of channel B are disconnected, and the
binary pump is rebooted, the error message is generated.
Probable cause Suggested actions
1 Damper disconnected.
Please contact your Agilent service
representative.
2 Defective damper.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Active-inlet valve and pump encoder of
channel B disconnected.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals