1260 Infinity Binary Pump User Manual 177
Maintenance
10
Exchanging the Solvent Selection Valve
Exchanging the Solvent Selection Valve
When If leaking internally (crossflow between the ports), or if one of the channels is blocked
Tools required p/n Description
8710-0899 Screwdriver, Pozidriv #1
Parts required p/n Description
G1381-60000 Solvent Selection Valve Upgrade Kit
Preparations Switch off the pump at the main power switch
1 Lift solvent bottles out of the solvent cabinet and place
them on the table. Disconnect the solvent tubes from the
solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.
2 Disconnect all tubings from the solvent selection valve.