178 1260 Infinity Binary Pump User Manual
10 Maintenance
Exchanging the Solvent Selection Valve
3
Using a Pozidriv screwdriver #1 loosen the holding
screws of the valve holder.
4 Carefully pull the valve holder out and disconnect the
valve cable at the connector.
5 Exchange the defective solvent selection valve. 6 Tighten the screws of the valve holder.