EasyManua.ls Logo

Agilent Technologies 1260 Infinity - Leak Rate Test

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1260 Infinity Binary Pump User Manual 137
Test Functions and Calibration
8
Leak Rate Test
Leak Rate Test
Introduction
The Leak Rate Test is used for verifying the internal tightness of the pump
and helps identifying parts which may have caused a leak.
System requirements
Minimum software revisions:
Lab Advisor B.02.03
Minimum firmware revisions:
A.06.50
NOTE
This test does not work in emulation mode, that is when the G1312B Binary Pump emulates
the G1312A Binary Pump. In case of an emulated module, convert to the original type first
(see “Replacing Module Firmware” on page 181).