110 1260 Infinity Binary Pump User Manual
7 Error Information
Module Error Messages
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
Pressure Signal Missing
Error ID: 2016
The pressure signal of the damper is missing.
The pressure signal of the damper must be within a specific voltage range.
If the pressure signal is missing, the processor detects a voltage of
approximately - 120 mV across the damper connector.
Probable cause Suggested actions
1 Lower pressure limit set too high.
Ensure the lower pressure limit is set to a value
suitable for the analysis.
2 Air bubbles in the mobile phase.
• Make sure that the degasser is in flow path
and works correctly. Purge the module.
• Ensure solvent inlet filters are not blocked.
3 Leak.
• Inspect the pump head, capillaries and
fittings for signs of a leak.
• Purge the module. Run a pressure test to
determine whether the seals or other
module components are defective.
4 Defective damper.
Please contact your Agilent service
representative.
5 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Damper disconnected.
Please contact your Agilent service
representative.
2 Defective damper.
Please contact your Agilent service
representative.