110 1260 Infinity Binary Pump User Manual
7 Error Information
Module Error Messages
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
Pressure Signal Missing
Error ID: 2016
The pressure signal of the damper is missing.
The pressure signal of the damper must be within a specific voltage range.
If the pressure signal is missing, the processor detects a voltage of
approximately - 120 mV across the damper connector.
Probable cause Suggested actions
1 Lower pressure limit set too high.
Ensure the lower pressure limit is set to a value
suitable for the analysis.
2 Air bubbles in the mobile phase.
Make sure that the degasser is in flow path
and works correctly. Purge the module.
Ensure solvent inlet filters are not blocked.
3 Leak.
Inspect the pump head, capillaries and
fittings for signs of a leak.
Purge the module. Run a pressure test to
determine whether the seals or other
module components are defective.
4 Defective damper.
Please contact your Agilent service
representative.
5 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Damper disconnected.
Please contact your Agilent service
representative.
2 Defective damper.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals