1260 Infinity Binary Pump User Manual 11
Introduction to the Binary Pump
1
Instrument and Operation
Principle of Operation
The binary pump is based on a two- channel, dual- piston in- series design
which comprises all essential functions that a solvent delivery system has
to fulfill. Metering of solvent and delivery to the high- pressure side are
performed by two pump assemblies which can generate pressure up to
600 bar.
Each channel comprises a pump assembly including pump drive, pump
head, active inlet valve with replaceable cartridge, and outlet valve. The
two channels are fed into a low- volume mixing chamber which is
connected via a restriction capillary coil to a damping unit and a mixer. A
pressure sensor monitors the pump pressure. A purge valve with
integrated PTFE frit is fitted to the pump outlet for convenient priming of
the pumping system.
Figure 2 The Hydraulic Path of the Binary Pump with Damper and Mixer
Ejbe]ZVY6 Ejbe]ZVY7
E^hidch
E^hidch
HZVah
HZVah
[gdb
WdiiaZ6
hdakZci
>caZikVakZ
B^m^c\
8]VbWZg
DjiaZi
kVakZ
DjiaZi
kVakZ
>caZikVakZ
idlVhiZ
EjbedjiaZi
Ejg\ZkVakZ
B^mZg
9VbeZg
[gdb
hdakZci
WdiiaZ7
EgZhhjgZ
hZchdg

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals