1260 Infinity Binary Pump User Manual 123
Error Information
7
Module Error Messages
Initialization Failed
Error ID: 2207, 2217
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time
window.
A maximum time is assigned for the complete pump- initialization cycle. If
the time is exceeded before initialization is complete, the error message is
generated.
Probable cause Suggested actions
1 Blocked active inlet valve.
Exchange the active inlet valve.
2 Defective pump drive assembly.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.