EasyManua.ls Logo

Agilent Technologies 1260 Infinity - Electronic Fuse of SSV Open; AIV Fuse

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1260 Infinity Binary Pump User Manual 113
Error Information
7
Module Error Messages
Electronic Fuse of SSV Open
Error ID: 2049
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent- selection valves in the module has drawn excessive
current causing the selection- valve electronic fuse to open.
AIV Fuse
Error ID: 2044
Inlet- Valve Fuse 0: Pump channel A
Inlet- Valve Fuse 1: Pump channel B
One of the active- inlet valves in the module has drawn excessive current
causing the inlet- valve electronic fuse to open.
Probable cause Suggested actions
1 Defective solvent selection valve.
Restart the pump. If the error message appears
again, exchange the solvent selection valve.
2 Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective active inlet valve.
Restart the module. If the error message
appears again, exchange the active inlet valve.
2 Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Related product manuals