1260 Infinity Binary Pump User Manual 113
Error Information
7
Module Error Messages
Electronic Fuse of SSV Open
Error ID: 2049
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent- selection valves in the module has drawn excessive
current causing the selection- valve electronic fuse to open.
AIV Fuse
Error ID: 2044
Inlet- Valve Fuse 0: Pump channel A
Inlet- Valve Fuse 1: Pump channel B
One of the active- inlet valves in the module has drawn excessive current
causing the inlet- valve electronic fuse to open.
Probable cause Suggested actions
1 Defective solvent selection valve.
Restart the pump. If the error message appears
again, exchange the solvent selection valve.
2 Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective active inlet valve.
Restart the module. If the error message
appears again, exchange the active inlet valve.
2 Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.