EasyManuals Logo

Agilent Technologies 1260 Infinity User Manual

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
114 1260 Infinity Binary Pump User Manual
7 Error Information
Module Error Messages
Temperature Out of Range
Error ID: 2517
Temperature Out of Range 0: Pump channel A
Temperature Out of Range 1: Pump channel B
One of the temperature sensor readings in the motor- drive circuit are out
of range.
The values supplied to the ADC by the hybrid sensors must be between
0.5 V and 4.3 V. If the values are outside this range, the error message is
generated.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.