1260 Infinity Binary Pump User Manual 115
Error Information
7
Module Error Messages
Temperature Limit Exceeded
Error ID: 2517
Temperature Limit Exceeded 0: Pump channel A
Temperature Limit Exceeded 1: Pump channel B
The temperature of one of the motor- drive circuits is too high.
The processor continually monitors the temperature of the drive circuits
on the main board. If excessive current is being drawn for long periods,
the temperature of the circuits increases. If the temperature exceeds the
upper limit, the error message is generated.
Probable cause Suggested actions
1 High friction (partial mechanical blockage)
in the pump drive assembly.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2 Partial blockage of the flowpath in front of
the damper.
Ensure the outlet valve is not blocked.
3 Defective pump drive assembly.
Please contact your Agilent service
representative.
4 Defective main board.
Please contact your Agilent service
representative.