1260 Infinity Binary Pump User Manual 65
Using the Pump
4
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface
Pump Control
The pump can be switched between following states: On, Off or to Standby.
In Standby, the pump motor is still controlled. When the pump is switched
on from standby, it does not re- initialize.
The optional seal wash pump can be controlled by either switching it off,
using it for a single time or specifying frequency and duration of periodic
wash intervals.
CAUTION
Upon initialization, the pump ignores the Maximum Flow Gradient value.
This can result in a rapid and uncontrolled pressure increase.
To prevent harm to the column, open the purge valve until the initialization is
finished.