1260 Infinity Binary Pump User Manual 173
Maintenance
10
Exchanging the Outlet Valve
Exchanging the Outlet Valve
1 Using a ΒΌ inch wrench disconnect the absorber capillary from the
outlet valve.
2 Unscrew the valve with the 14 mm wrench and remove it from the
pump body.
3 Do not disassemble the outlet valve, as this can damage the valve.
4 Reinstall the outlet valve and tighten it using a torque wrench (12 Nm).
When if leaking internally
Tools required Description
Wrench, 1/4 - 1/5 inch
Wrench 1/4 inch
Wrench, 14 mm
Parts required p/n Description
G1312-60067 Outlet valve 1220/1260
Preparations Switch off the pump at the main power switch