EasyManua.ls Logo

Agilent Technologies 1260 Infinity - Exchanging the Outlet Valve

Agilent Technologies 1260 Infinity
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1260 Infinity Binary Pump User Manual 173
Maintenance
10
Exchanging the Outlet Valve
Exchanging the Outlet Valve
1 Using a ¼ inch wrench disconnect the absorber capillary from the
outlet valve.
2 Unscrew the valve with the 14 mm wrench and remove it from the
pump body.
3 Do not disassemble the outlet valve, as this can damage the valve.
4 Reinstall the outlet valve and tighten it using a torque wrench (12 Nm).
When if leaking internally
Tools required Description
Wrench, 1/4 - 1/5 inch
Wrench 1/4 inch
Wrench, 14 mm
Parts required p/n Description
G1312-60067 Outlet valve 1220/1260
Preparations Switch off the pump at the main power switch

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Related product manuals