172 1260 Infinity Binary Pump User Manual
10 Maintenance
Exchanging the Active Inlet Valve (AIV) or its Cartridge
6 Push the cartridge into the new active inlet valve.
7 Screw the new valve into the pump head. With the 14 mm wrench turn
the nut until it is hand tight.
8 Position the valve so that the solvent inlet tube connection points
towards the front.
9 Using the 14 mm wrench tighten the nut by turning the valve in its
final position (not more than a quarter turn). Do not overtighten the
valve.
10 Reconnect the Active Inlet Valve cable to the connector in the Z- panel
and the inlet tube to the valve.
11 Reinstall the front cover.
6Xi^kZ^caZikVakZXVgig^Y\Z
6Xi^kZ^caZikVakZ
After an exchange of the valve it may be required to pump several mL of the solvent used in
the current application before the flow stabilizes at a pressure ripple as low as it used to be
when the system was still working properly.