12 1260 Infinity Binary Pump User Manual
1 Introduction to the Binary Pump
Instrument and Operation
Damper and mixer can be bypassed for lowest delay volume of the binary
pump. This configuration is recommended for low flow rate applications
with steep gradients, see the Agilent 1260 Infinity Binary LC System
Manual.
Figure 3 on page 12 illustrates the flow path in low delay volume mode.
For instructions on how to change between the two configurations, see
“Convert the Binary Pump to Low Delay Volume Mode” on page 85.
Figure 3 The Hydraulic Path of the Binary Pump with Bypassed Damper and Mixer
For pump specifications, see “Performance Specifications” on page 26.
Bypassing the mixer while the damper remains in line is not a supported configuration and
may lead to undesired behavior of the binary pump.
EjbedjiaZi
idlVhiZ
B^mZg
9VbeZg
Ejbe]ZVY6 Ejbe]ZVY7
DjiaZikVakZ
>caZikVakZ
HZVah
HZVah
E^hidch
E^hidch
B^m^c\X]VbWZg
Ejg\ZkVakZ
[gdbhdakZci
WdiiaZ7
>caZi
kVakZ
[gdbhdakZci
WdiiaZ6
DjiaZi
kVakZ
EgZhhjgZ
hZchdg