Chapter 4 Angular Measurements
Angular Position Measurements
4-46 Measurements Reference Guide
CAUTION Never use an ordinary nut in place of the clutch breakaway nut. The
ordinary nut is not designed to provide the breakaway-under-load
protection provided by the Agilent-provided nut.
After you have assembled and tightened the linkage, unscrew the clutch
breakaway nut and position it about 5 mm (1/4-inch) up the shaft, so it
does not interfere with operation of the clutch.
Spherical washer sets are particularly useful in right-angle situations,
when used with the Agilent-provided adapter, or on complex linkages
where out-of-squareness can be a problem.
Figure 4-25. Clutch
1
Clutch control shaft
2 Clutch breakaway nut
3 Clutch washer
4 Top of measurement fixture
*
**
***
“Neutral” position
Side force rocking (jogging) clutch
Breakaway nut tightened to lock clutch in its engaged position.
This is done for setups only. Do not lock the clutch during measurements.