Chapter 7 Squareness Measurements in a Vertical Plane
Mounting and Aligning Optics for the Second Axis Measurement
Measurements Reference Guide 7-13
Mounting and Aligning Optics for the Second
Axis Measurement
This section explains how to mount and align the optics for the second axis
measurement.
Mounting the optics
To mount the optics for the second axis measurement, follow these steps:
1 Move the moveable part of the machine to its start position.
2 Remove the current interferometer assembly from the spindle and
disassemble the assembly.
3 Attach the turning mirror base to the optical square using the turning
mirror attachment screw (Figure 7-7).