Chapter 4 Angular Measurements
Angular Position Measurements
4-70 Measurements Reference Guide
Figure 4-34. Replacing the clutch washer
Step Art Procedure
1—Remove optic, if necessary.
21Remove four M4 cap screws holding retainer ring to fixture top plate.
31Remove clutch control shaft from retainer ring.
42Remove clutch washer from base of clutch shaft.
53Cut clutch washer and remove from shaft.
64Slide new clutch washer onto shaft—paper-covered adhesive side first.
7—Remove protective paper from adhesive.
8—Press washer onto base.
91Replace shaft in retainer ring.
10 1 Fasten retainer ring to fixture top plate.
1
234