Chapter 2 Linear Measurements
Setting Up for the Measurement
Measurements Reference Guide 2-5
Figure 2-3. Set Up Environmental Compensation screen
The system uses information entered on this screen to compensate for
measurement inaccuracies caused by environmental factors such as air
temperature.
If you are using sensors, the system automatically compensates for
measurement inaccuracies caused by environmental conditions. Go
directly to “Placing and Using the Sensors.”
If you are not using sensors, you must manually enter values on this
screen that will compensate for measurement inaccuracies. Continue with
step 8.
8 On the Set Up Environmental Compensation screen, do the following:
a. In the Environmental Compensation box, select Manual.