Chapter 2 Linear Measurements
Making the Measurement
Measurements Reference Guide 2-29
8 Continue moving the machine’s moveable part to each point and recording
measurements until the machine has reached the last measurement
position.
NOTE If at any point the Reset Laser message flashes, the laser beam path has
been broken. Move the retroreflector back to the last position measured
before the beam was broken, then select Reset Laser to reset the system.
After you record the last measurement, the Analyze Data: LINEAR screen
is displayed (Figure 2-17).
Figure 2-17. Analyze Data: LINEAR screen
9 On the Analyze Data: LINEAR screen, select Save Data to create or
update a data file with the measurement data.
Refer to the Agilent 5530 Getting Started Guide and online help for more
information about using this screen. Otherwise, you are finished making
the measurement.