Introduction
Data rates
Released: March 2004 6850 Series II GC User Information page 20 of 256
LAN card
The LAN card in the GC conforms to industry standards and enables
communication with the Cerity Chemical and ChemStation products as well as
other external computers equipped with a LAN interface and suitable software.
If you are using Cerity Chemical or ChemStation, the method of setting the IP
address depends on the type of LAN card you have installed in your GC. See
“Setting your IP address” on page 245 for more information.
Data rates
Digital output to the ChemStation/Cerity Chemical is available at 11 speeds
ranging from 0.1 Hz to 200 Hz, capable of handling peaks from 0.001 to
2 minutes wide. The data rate is selected from the ChemStation/Cerity Chemical
and is described in the ChemStation/Cerity Chemical documentation.
Keypad lockout
While the ChemStation/Cerity Chemical is controlling the GC, you cannot load
methods from the GC keypad. The ChemStation/Cerity Chemical can also
disable other keys and functions, for example, what information is available on
the scrolling display.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals