Warning messages
WARNING 100 Oven sensor missing
Possible causes: Sensor not connected
Sensor defective
The GC main board could not detect the oven temperature sensor. To prevent
a thermal runaway, the oven is set to off and the instrument is not ready.
WARNING 101 Excess heater power
Possible causes: Heater wattage total >210W
Defective main board
The total wattage for the detector, inlet, and auxiliary heated zones exceeds
210 Watts. This value is only checked at power on. The three heated zones
are set to ’not installed.’
WARNING 103 Signal 1 buffer full
Possible cause: Data collection device is off-line and signal buffer is full
If your integrator or ChemStation goes off line during a run, signal data is
stored in the GC’s signal buffer. When the buffer is full, the warning appears
and no new data is stored.
WARNING 104 Signal 2 buffer full
Possible cause: Control module is off-line and its signal buffer is full
If the control module goes off line during a run, signal data is stored in the
GC’s signal buffer. When the buffer is full, the warning appears and no new
data is stored.
WARNING 105 Analog out data loss
Possible causes: Defective main board
Bad power supply (less likely)
The GC analog processors are not operating correctly. Some analog data is
lost, but digital data may be uncorrupted.
WARNING 106 Data loss
Possible causes: Defective main board
Bad power supply (less likely)
The GC digital processors are not operating correctly. Both analog and digital
data are lost.