Sequence warning messages
Released: March 2004 6850 Series II GC User Information page 86 of 256
WARNING 146 Memory reset
Possible causes: The main board battery failed
The GC firmware was updated
All GC methods, the run log, and the log book are stored in battery-backed GC
memory. If the backup battery fails and power is turned off, this data is lost
and the warning appears when the GC is turned on. If you update the GC
firmware, this data is erased during the update, and the warning appears after
restart.
WARNING 901 GC firmware
Possible cause: GC firmware revision is A.01.xx and a purged packed inlet is installed
A purged packed inlet is detected but the current GC firmware is out of date
and may not support the new inlet. Contact Agilent Technologies to upgrade
your GC firmware. Version A.02.00 or higher is required.
Sequence warning messages
Autoinject aborted
Possible cause: The injector sequence was aborted during a run or injection. Other messages detail why.
Bottle # out of range
Possible causes: Incorrect last vial parameter
Wrong turret installed for your method
Turret was changed but not aligned
The GC failed to find the vial specified in the method because that vial number
is not available in the turret installed. This message may appear if you changed
the turret type but did not perform the alignment procedure.
Incomplete injection
Possible causes: Bent syringe needle
Defective plunger or syringe carriage
The injector syringe carriage did not reach the position at which the needle is
fully inserted in the inlet. The injection could not be completed or some sample
may have escaped.
Inj door or mounting
Possible causes: Injector door open
Injector not mounted properly on GC
Defective safety switch on injector
During an injection routine, the injector halts for safety reasons if either the
injector door is open or if the injector is not properly mounted on the GC post.
Injector comm error
Possible cause: Defective injector board
During an injection in a sequence, the GC lost communication with the injector.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals